Strany 1 - LATITUDE

OPERATOR’S MANUALZOOM®LATITUDE®Programmer/Recorder/MonitorModel 3120CAUTION: Federal law(USA) restricts this deviceto sale by or on the orderof a p hy

Strany 2

External PrinterYou can use an external printer if i t is supported by the pulse generatorsoftware applicatio n. Use only compatible external printers

Strany 3 - Table of Contents

NOTE: The Model 6577 Telemetry Wand has been tested for 25 sterilizationcycles and more than this number o f sterilization cycles is not recommended.D

Strany 4

[1] On/Off button [2] External printer connector [3] USB ports [4] External VGA monitor connectorFigure 2. Left side panel of the PRM• Connect the Mod

Strany 5 - WARNINGS

• Use the equipotential stud connection to equa lize the PRM’s galvan icpotential with other electrical equipment. Using this terminal toprovide equal

Strany 6 - PRECAUTIONS

[1] AntennaFigure 4. Antenna in the upright positionUSING THE PRMStartup ScreenThe PRM has a touchscreen and a tethered stylus allowing you to select

Strany 7 - System Features

Figure 5. Startup screenWhen the PRM is powered On, the startup screen is displayed and containsthe foll owi ng information:• The ECG Display, which s

Strany 8

Changing Parameter ValuesThe screens for man y of the features contain parameter information that canbe changed via either a palette window or a keybo

Strany 9 - Optional External Equipment

Copy ButtonOn those screens that contain a Copy button, you can simply copy parametervalues from one screen to another. S elect the Copy button. A win

Strany 10 - PREPARING THE PRM FOR USE

NOTE: The values as s et up on the startup screen will be the de faults usedfor the application traces. The corresp onding values ca n be chang ed fro

Strany 11

PRMharddrive.DatasavedtotheharddrivecanthenbeexportedtotheUSBpen drive through the Export Data feature of the Patient Data Managementutility, accessib

Strany 13

When using a non -Boston Scientific computer (e.g., a clinic PC), enteryour password to access encryptedpatientdatastoredontheUSBpen drive.NOTE: The US

Strany 14 - Startup Screen

2. Select the USB Drive or Programmer option to indicate the location fromwhich you want to prin t patient records.3. Select the p atient records you

Strany 15 - Figure 5. Startup screen

5. When the delete operation initiates successfully, the system displays amessage stating that Protected Health In formation is being deleted fromthe

Strany 16 - Figure 7. Keyboard w

Utilities ButtonIf desired, before accessing the pulsegenerator software application, you canselect the Utilities button to perform the following acti

Strany 17 - ECG Display

a. To become familiar with the software without interrogating a pulsegenerator, select the DEMO button; the main ap pli cation screen willappear and t

Strany 18 - Quick Start Button

[1] STAT PACE [2] DIVERT THERAPY [3] STAT SHOCK [4] PROGRAM [5] INTERROGATEFigure 8. Right-side key padThe following description of the right-side key

Strany 19

• [2] Press the paper-feed key to scroll the printer paper on the internalprinter/recorder.• [3] Press the ca libration key to cause the internal prin

Strany 20

permanence of a patient’s record, store the printed paper away from directsunlight, heat, or fumes from organic compound s. Storage temperaturesabove

Strany 21

[1] Write-protect tab closed (black tab covering hole) [2] Sliding shuttleFigure 10. Patient Data DiskThediskmustbeinsertedwiththearrow o n the top le

Strany 22

• Do not attach paper clips, staples, or rubber bands to disks.• Do not try to open the sliding shutter that covers the disks (Figure 10 onpage 24).•

Strany 23 - Select PG Button

Table of ContentsINFORMATION FOR USE ... 1Trademark Statement ...

Strany 24 - Indicator Lights

CAUTION: The PRM and ZWT are not waterp roof or explosion-proof andcannot be sterilized. Do not use them in the presenc e of flammabl e gasmixtures inc

Strany 25 - Figure 9. Left-side keypad

ServiceFor questions regarding operation or repair of the PRM or ZWT, contact BostonScientific using the information on the back cover. The PRM and ZWT

Strany 26 - Thermal Paper Storage

Table 2. Possible causes and correc tive actions fo r PRM pro blems (continued)Symptom Possible Cause Corrective Actio nInternalprinter/recorder: nopr

Strany 27 - Patient Data Disk

Table 2. Possible causes and corrective actions for PRM problems (cont inued)Symptom Possible Cause Corrective ActionIncorrect telemetrywandUse only t

Strany 28 - Figure 10. Patient Data Disk

Table 3. Possible causes and correctiv e actions for ZWT problemsSymptom Possible Cause Corrective Actio nUSB cable notsecurely connectedto ZWTRemove

Strany 29 - Operation and Sto rage

Figure 12. External ECG Monitor ConfigurationTo di splay a tracing on a n external ECG monitor and the P RM, set upequipment as shown above (Figure 12

Strany 30

Table 4. Symbols on packaging (continued)SymbolDescriptionLot numberDate of manufactureNon-ionizing electromagneticradiation; ZIP telemetry indicatorl

Strany 31 - Tr oubleshooting

Table 4. Symbols on packaging (continued)SymbolDescriptionDefibrillation-proof type CF appliedpartDefibrillation-proof type BF appliedpartECG cable conn

Strany 32

Table 4. Symbols on packaging (continued)SymbolDescriptionOn indicator lightAssembly numberThis side upFragile, handle with careKeep dryDo not use hoo

Strany 33

Electromagnetic Compatibility StandardsThePRMhasbeentestedandfoundtocomply with the applicable portions ofthe electromagnetic compatibility (EMC) stan

Strany 34

Specifications... 40

Strany 35 - Symbols on Packaging

Table 5. Guidance and manufacturer’s declaration - electromagnetic emissions -for all eq uipment and systemsEmissions testComplianceElectromagneticenv

Strany 36

Table 6. Guidance and manufacturer’s declaration - electromagnetic immunity -for all equipment systems (continued)Immunity testIEC 60601 testlevelComp

Strany 37

Table 7. Guidance and manufacturer’s declaration - electromagnetic immu nity -for equipment and systems that are not life supportingImmunity testIEC 6

Strany 38 - Compliance Standards

Table 7. Guidance and manufacturer’s declaration - electromagnetic immunity -for equipment and systems that are no t life supporting (continued)Immuni

Strany 39 - Federal Communications Commi

Table 8. Recommended separation distances between portable and m obile R Fcommunications equipment, and the PRM / ZWT (continued)Separation distance a

Strany 40

Table 9. PRM Nominal Specifications (continued)CharacteristicNominalTransport and storage humidity25% to 95%Operating altitude≤ 2000 mTransport and sto

Strany 41 - The customer or th

Table 9. PRM Nom inal S pecifications (continued)CharacteristicNominalBandwidth < 1 MHzModulationASK/OOKEffective radiated power-1.3 dBmInternal pri

Strany 42

Table 10. ZWT No minal Sp ecifications (continued)CharacteristicNominalWeight (approximate ) 0.6 kg ( 1.3 lb)Power rating5VDCPower cordPower supplied v

Strany 45 - to lerance

INFORMATION FOR USETrademark StatementThe following are t rademarks of Boston Scientificoritsaffiliates: LATITUDE,Quick Start, ZIP, ZOOM.Description and

Strany 48

ManufacturerBoston Scientific4100 Hamline Avenue NorthSt. Paul, MN 55112-5798 USAwww.bostonscientific.com1.800.CARDIAC (227.3422)+1.651.582.4000© 2014 B

Strany 49

• Special committee on radio interference. Other equipment mayinterfere with the PRM and ZWT, even if that equipment comp lies with theInternational S

Strany 50

Preparations for Use• Telemetry wand shipped nonsterile. The Model 6577 Telemetry Wandis shipped nonsterile. If the telemetry wand is to be used in a

Strany 51

The PRM also:• Provides a direct interface between anexternal stimulator and animplanted pulse generator for programmed electrical s timul ation (PES)

Strany 52 - *357434-009*

•ZIP™telemetry, a cordless, hand s-free radio frequency (RF)communication option that allows the PRM to communicate with the pulsegeneratorNOTE: To co